Recall Date: 07/31/2009 |
Title: Physio-Control Inc., LIFEPAK CR Plus Automated External Defibrillators (AEDs) |
Units: Physio-Control LIFEPAK CR PLUS AEDs with the serial numbers below are affected by this recall. Serial Numbers: 37026963, 37026983, 37026984, 37026997, 37027002,37027008, 37027039, 37027040, 37027049, 37027053, 37027063, 37027065, 37027066, 37027070, 37027071, 37027073, 37027075, 37027090, 37027099, 37027105 37027122, 37027197, 37027529, 37027569, 37031393, 37037850, 37037893, 37037986, 37038002, 37038211, 37038365, 37135986, 37154526, 37154638 |
Issue: An extremely humid environment may cause the LIFEPAK CR Plus AED to improperly analyze the rhythm correctly and may cause the device to delay or fail to delivery therapy. |
Fix: Contact Physio-Control Customer Care 1-800-442-1142, 6 AM through 4 PM Pacific Time |